Nuclear Science and Techniques

《核技术》(英文版) ISSN 1001-8042 CN 31-1559/TL     2019 Impact factor 1.556

Nuclear Science and Techniques ›› 2008, Vol. 19 ›› Issue (5): 297 doi: 10.1016/S1001-8042(09)60007-5

• ADVANCED NUCLEAR INSTRUMENTATION AND DETECTION • Previous Articles     Next Articles

A compact gamma camera with scintillation array and parallel-hole collimator

ZHU Jie1, 2,* MA Hongguang1 MA Wenyan1 ZENG Hui2 WANG Zhaomin2 XU Zizhong2   

  1. 1 The Second Artillery Engineering Institute, Xi’an 710025, China
    2 Department of Modern Physics, University of Science and Technology of China, Hefei 230026, China
ZHU Jie MA Hongguang MA Wenyan ZENG Hui WANG Zhaomin XU Zizhong. A compact gamma camera with scintillation array and parallel-hole collimator.Nuclear Science and Techniques, 2008, 19(5): 297     doi: 10.1016/S1001-8042(09)60007-5

Abstract: A new compact gamma camera for small object imaging has been developed. It consists of a pixelized Nal(Tl) scintillator array coupled to a position sensitive photomultiplier tube (Hamamatsu R2486) with a parallel-hole lead collimator. The compact camera has better spatial resolution than Anger camera. The average value of intrinsic spatial resolutions is 2.3 mm (FWHM). The overall spatial resolution (FWHM) is 3, 5 and 6 mm at 0, 2.5 and 3 mm SCD (source-to-collimator distance), respectively. The phantom studies with the compact camera have demonstrated that parallel-hole collimator gamma camera is a practical technique for nuclear medicine application.

Key words: Compact gamma camera, Parallel-hole collimator, Scintillation array, PSPMT