Abstract:

The extraction algorithms for pulse amplitude and smoothing of energy spectrum have a great influence on energy spectrum of γ-rays during the digital detection and analysis procedure. For a CdZnTe digital detector system, different extraction algorithms for pulse amplitude and smoothing of energy spectrum are discussed in this paper. The results show that extraction of pulse amplitude using the first-order derivative method and smoothing of energy spectrum using the wavelet transformation method may obtain energy spectrum with good performance.

Key words: Pulse amplitude extraction, Energy spectrum smoothing, Optimization algorithm