Calibration method for electrode gains in an axially symmetric stripline BPM

WU Fang-Fang,1 ZHOU Ze-Ran,1 YANG Yong-Liang,1 SUN Bao-Gen,1 ZOU Jun-Yin,1 LU Ping,1 CHENG Chao-Cai,1 LUO Qing, XU Jun 1   

  1. 1National Synchrotron Radiation Laboratory & School of Nuclear Science and Technology, University of Science and Technology of China, Hefei 230029, China
  • Contact: SUN Bao-Gen E-mail:bgsun@ustc.edu.cn
  • Supported by:

    Supported by the Natural Science Foundation of China (Nos. 11175173, 11375178 and 11005105)

PDF ShareIt Export Citation
WU Fang-Fang, ZHOU Ze-Ran, YANG Yong-Liang, SUN Bao-Gen, ZOU Jun-Yin, LU Ping, CHENG Chao-Cai, LUO Qing, XU Jun . Calibration method for electrode gains in an axially symmetric stripline BPM.Nuclear Science and Techniques, 2014, 25(5): 050102     doi: 10.13538/j.1001-8042/nst.25.050102

Abstract:

The four electrodes in the stripline beam position monitor (BPM) for Hefei Light Source (HLS II) storage ring are of axially symmetric type. We have derived a new calibration method of electrode gains for this type stripline BPM. The gain fit error of different data grids was analyzed, and the 5mm by 5mm grid is the best. The electrode gains of two stripline BPMs (HLS II SR-BD-STLB1 and HLS II SR-BD-STLB2) were obtained based on offline calibrated data. The results show that data after fitting gains are improved, with the electrode gains being between 0.94 and 1.15.

Key words: Stripline beam position monitor, Electrode gain, Axially symmetric, Calibration