An efficient method and system for simultaneously measuring the performance of vast scintillating materials

CHEN Yan1,2,3 ZHANG Zhiming1,2 LI Daowu1,2 SHUAI Lei1,2 HUANG Xianchao1,2 TANG Haohui1,2 LI Ting1,2 ZHUANG Kai1,2 WANG Yingjie1,2,3 LIU Yantao1,2,3 ZHANG Yiwen1,2,3 CHAI Pei1 LIU Shuangquan1,2 ZHU Meiling1,2 WANG Baoyi1,2 WEI Long1,2,*   

  1. 1Laboratory of Nuclear Analysis Techniques,Institute of High Energy Physics,Chinese Academy of Sciences,Beijing 100049,China 2Beijing Engineering Research Center of Radiographic Techniquesand Equipment,Beijing 100049,China 3Graduate University of Chinese Academy of Sciences,Beijing 100049,China
CHEN Yan, ZHANG Zhiming, LI Daowu, SHUAI Lei, HUANG Xianchao, TANG Haohui, LI Ting, ZHUANG Kai, WANG Yingjie, LIU Yantao, ZHANG Yiwen, CHAI Pei, LIU Shuangquan, ZHU Meiling, WANG Baoyi, WEI Long . An efficient method and system for simultaneously measuring the performance of vast scintillating materials.Nuclear Science and Techniques, 2012, 23(3): 144-149     doi: 10.13538/j.1001-8042/nst.23.144-149

Abstract:

A method and system for automatically and simultaneously measuring the light output of multiple scintillators,or each scintillating unit of an array,were developed.Using a large area flat panel PSPMT H8500,the light output and energy resolution were obtained automatically by comparing with reference scintillators or array using an Energy Table,Look-up Table and energy spectrum data.The aim of developing an efficient performance evaluation of scintillators was achieved.Using the method,a scintillator performance testing system was set up and six LYSO crystals and a 3×3 LYSO array were measured.The results showed that the light output and energy resolution were accurately measured automatically.The deviation of repeat measurements for the same sample was not more than 2%,and the nonlinear deviation of the system was not more than 3%.The system is suitable for measuring the performance of crystals,especially where the mass measurement of crystals and arrays is required.

Key words: Scintillating materials, Light output, Energy resolution, Measurement