A mathematical method to calculate efficiency of BF3 detectors

SI Fenni* HU Qingyuan PENG Taiping   

  1. Institute of Nuclear Physics and Chemistry, China Academy of Engineering Physics, Mianyang 621900, China
PDF ShareIt Export Citation
SI Fenni, HU Qingyuan, PENG Taiping. A mathematical method to calculate efficiency of BF3 detectors.Nuclear Science and Techniques, 2009, 20(3): 181-183     doi: 10.13538/j.1001-8042/nst.20.181-183

Abstract:

In order to calculate absolute efficiency of the BF3 detector, MCNP/4C code is applied to calculate relative efficiency of the BF3 detector first, and then absolute efficiency is figured out through mathematical techniques. Finally an energy response curve of the BF3 detector for 1~20 MeV neutrons is derived. It turns out that efficiency of BF3 detector are relatively uniform for 2~16 MeV neutrons.

Key words: BF3 detector, Absolute efficiency, Relative efficiency, Energy response curve