Optimization of OSL measurement conditions for resistance element

GUO Jingyuan1 TANG Qiang1 JIA Yuxin2 LIU Xiaowei1 LIU Yanbing2   

  1. 1(School of Physics and Engineering, Sun Yat-sen University, Guangzhou 510275, China) 2(Key Laboratory of Occupational Disease Prevention and Treatment, Occupation Disease Prevention and Treatment Center of Guangdong Province, Guangzhou 510300, China)
  • Received:2013-11-29 Revised:2014-01-12 Online:2014-05-10 Published:2014-05-08
  • Supported by:

    the National Natural Science Foundation of China

Abstract:

Background: Dosimetry material used for the evaluation of radiation accident dose is an important basis for the reconstruction of accident dose. It is of great significance to select materials for dosimeters suitable for the emergency and accident site to evaluate absorbed dose. Purpose: SMD resistors which are widely used in the circuit components were chosen, and the optimized optically stimulated luminescence (OSL) measurement conditions were studied. Methods: Under different measurement conditions such as preheating temperature and bleach time, the linearity and stability of dose response were studied, and the optimized conditions were determined. Results: It was found that the linearity and stability of OSL dose response performed well in the dose range from 0.1 Gy to 10 Gy under the optimized conditions. Conclusion: It is a good choice to use the SMD resistor as a suitable dosimeter for the accident dose reconstruction.

Key words: Optically stimulated luminescence (OSL), SMD resistor, Dose response, Measurement conditions